Puritan Bennett Product Detail
Model: GoodKnight 425
Bilevel is the generic term to describe these machines. BiPAP, for Bilevel Positive Airway Pressure, is registered with Respironics; ResMed uses VPAP for Variable Positive Airway Pressure.
The Puritan Bennett GoodKnight 425 Bilevel machine includes the following features:
- Smallest, lightest Bilevel, BiPAP or VPAP on the market.
- SilverLining 3.10 Data Management software option to download compliance and session data.
- International power supply to operate on any AC current from 110 to 240 without adjustment to the machine. (An adapter may be needed to plug the CPAP into a wall outlet outside the US.)
- Direct DC operation without the need for a DC/AC inverter, just a cable and adapter which are sold separately on our site.
- Automatic altitude compensation adjustment.
- Two washable, reusable pollen filters are included; optional ultra-fine filters are available.
- 2 year manufacturer warranty against defects.
Specifications:
- Performance:
- Performance range: 3 to 25 cm H2O (+/- 1 cm H2O)
- Pressure adjustment: 0.5 cm H2O
- Maximum pressure at the patient connection port under single fault condition: 36 cm H20
- Maximum sound level warranted at one meter: 31 dBA at 10 cm H2O
- Ramp time: 0 to 30 minutes (5 min. increments)
- Sound level measured in accordance with ISO 17510-1:2002: 25.5 dB(A)
- Physical Characteristics:
- Size: 5.6" W x 2.9" H x 7.7" L (14.2 cm x 7.5 cm x 19.5 cm)
- Weight: 1.55 lb (780 g)
- Power Characteristics:
- Power supply: Input range 100-240 V, 50-60Hz
- Power consumption: 40 W
- Battery: 12 VCC, power supply 13 VCC
- Environmental Requirements:
- Operating temperature: +41 F to +104 F (+5C to 40 C)
- Altitude range: 0 to approximately 7800 ft (0 to 2400 m)
- Operating humidity: Between 15% and 95% noncondensing
- Operating pressure range: 700 hPa to 1060 hPa
- Storage temperature: +4 F to +140 F (+20C to 60 C)
- Storage humidity: Up to 95% noncondensing
- Storage pressure range: 500 hPa to 1060 hPa
- Settings:
- Parameters: IPAP level
- Minimum Value: 3 cm H2O
- Maximum Value: 25 cm H2O
- Default Value: 15 cm H2O
- Step: 0.5 cm H20
- Parameters: EPAP level
- Minimum Value: 3 cm H2O
- Maximum Value: IPAP or 20 cm H2O
- Default Value: 5 cm H2O
- Step: 0.5 cm H20
- Parameters: Ramp time
- Minimum Value: 0 min.
- Maximum Value: 30 min.
- Default Value: 15 min.
- Step: 5 min.
- Parameters: Starting ramp pressure
- Minimum Value: 3 cm H2O
- Maximum Value: EPAP
- Default Value: 3 cm H2O
- Step: 0.5 cm H20
- Parameters: Pressure rise time
- Minimum Value: 1 (~300 ms)
- Maximum Value: 5 (~600 ms)
- Default Value: 3 (~450 ms)
- Step: 1 (~125 ms)
- Parameters: Inspiratory sensitivity
- Minimum Value: 1
- Maximum Value: 10
- Default Value: 5
- Step: 1
- Parameters: Expiratory sensitivity
- Minimum Value: 1
- Maximum Value: 10
- Default Value: 5
- Step: 1
Request Info
|
|