The forced air circulation system permits stable
temperature control, uniform heat distribution, humidification,
and effective isolation of the infant from airborne
contaminants.
The unit provides a connection for the
administration of supplemental oxygen.
Power Requirements:
Model C300-2: 110/120V~; 50/60 Hz, 175/205 W.
Model C300-2E: 220/240V~; 50/60 Hz, 175/205W.
Chasis Leakage Current: 100 uA or less.
Alarms:
Sensor: actuates if high temperature alarm sensor is electrically open or shorted.
High Temperature: Activates if incubator display reads 38.5C +/- 0.5C.
Power Failure: Activates if main AC power fails and no DC back-up power is present.
Heater Temperature: Activates if heater temperature exceeds 77C (nominal).
Alarm Reset switch: Cancels Alarm indicators if alarm condition has been corrected (except power failure alarm).
Temperature Set Point Range: 22.0 +/- 0.5C to 38.0 +/- 0.1C
Temperature Rise Time: 45 minutes
Temperature Variability: 0.2C
Temperature Overshoot: 1.5C maximum
Correlation of Displayed Incubator Temperature to Actual Incubator Temperature at Temperature
Equilibrium: +/- 1 C.
Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium:
0.5C
Oxygen Concentration Range: 21% to at least 58 % at 6 LPM flow rate
Humidity: Typically between 50% and 70%
Air Velocity Over Mattress: Does not exceed < 20cm/sec (40 ft/min)
Ambient Operating Temperature: 21C to 28C.
Noise Level within Hood: 60 dBA maximum with 50 dBA or less ambient. 80 dBA maximum
during alarm.
Nominal Dimensions:
Without stand - Height: 60.3 cm (23.75 in) Width: 52.7 cm (20.75 in)
Length: 81.3 cm (32 in).
With stand - Height: 128.9 cm (50.75 in) Width: 55.9 cm (22 in), length: 81.3 cm (32 in).