Temperature rise time (ambient temperature 25°C): < 30 minutes
Temperature variability: < 1°C
Temperature uniformity (Mattress in lever): < 1.5°C
Correlation of displayed incubator temperature to actual incubator temperature at temperature equilibrium: <0.7°C
Skin sensor accuracy: <0.1°C
Alarms
Power Fail (Power Failure): Actuates if main AC power fails and no DC back-up power is present or switching to battery power.
Fan Alarms: Motor trouble or fan rotate speed less low or fan stop
DC Low Voltage Alarms: DC 12V or Battery supply low 10.5V or DC24V low 22.5V
Sensor Alarms: Actuates if there is a sensor failure or skin sensor isn't placed in right place
Temperature deviation alarm: Air temperature or skin temperature too high or air temperature or skin temperature too low.
Over-temperature alarm: Temperature too high in hood under air temperature control mode or skin temperature control mode.
E Rom trouble alarms: Within temp controller system trouble
Whole equipment physical dimensions
Length: 170 cm
Width: 50 cm
Incubator body weight (including 2 batteries: 80 kg
Stand weight: 25 kg
The max. force load of the IV Pole: 20N
The max. force load of the mattress: 98N
Mattress's length: 63 cm
Mattress's width: 35 cm
Other specifications
Noise level within the hood environment: <60dB(A)
With ambient of equal or less than 50Dba
Carbon Dioxide (CO2) Lever within the hood: Less than 0.5% when a 4% mixture of CO2 in air is delivered at 750 ml / min at a point 10 cm above the center of the mattress
Air flow above mattress: <0.35m/s
Incubator service life: Incubator's use life is limited, generally 8 years, but it is according to every incubator different using frequency, operational status and environment of working and storage. The service life will be changed.