Portable Multiparameter Patient Monitor Product Detail
Model: UM-P400
UM-P400 is mini sized (portable) and multiparameter patient monitor. The unit can measure such features as ECG, Pulse, SpO2, NIBP, Temperature, and Respiratory. With the Trend View function, monitor can monitor trends for 24 hours. For easy portability, the unit can be powered by internal battery as well as direct AC electric source. To best suit the needs of your facility, this monitor is available in the following configurations:
- UM-P300 - ECG/Resp/Pulse/Temp
- UM-P310 - ECG/Resp/Pulse/Temp/SpO2
- UM-P320 - ECG/Resp/Pulse/Temp/NIBP
- UM-P330 - SpO2/NIBP
- UM-P400 - ECG/Resp/Pulse/Temp/SpO2/NIBP
Specifications:
- General
-
2 Trace waveforms display
-
High resolution graphic LCD with CCFL backlight
-
Refined jog-dial control
-
Universal power voltage input
-
Memorized Data and waveform print
-
Data communication
-
ECG : Defibrillator/ESIS protected
-
SpO2 : Principle of operation is red/infrared absortion
-
NIBP :
-
One hose cuff
-
Stepdown air relealse with solenoid valve
-
Oscillometric method
-
Temperature : Compatible YSI probe
-
Respiration : Thoracic Impedance Detection with ECG Electrode
-
ECG
-
CMRR (Common Mode Rejection Ratio) : 100dB
-
Input : RA, LA, LL(3 Lead ECG Patient Cable)
-
Lead Selection : I, II, III, and CAL
-
Frequency Bandwidth : 0.5~40Hz
-
Gain Selection : 0.5, 1, 2mV/cm
-
Isolation Breakdown : Voltage >2500VAC 60Sec 60Hz
-
Leakage : Current <10uA
-
Notch Filter : 50/60 Hz
-
Defibrillator/ESIS : Protected
-
SpO2
-
Measuring : Range 30~100%
-
Accuracy :
- 1%(90~100%)
- 2%(80~89%)
- 3%(65~79%)
- 4%(below 65%)
-
Resolution : 1%
-
Alarm Limit :
-
NIBP
-
Range :
-
Adult : 20~300mm Hg
-
Neonate : 10~150mm Hg
-
Accuracy : ¡¾3mm Hg or 2%
-
Resolution : 1mmHg
-
Measuring Interval Time OFF : 1~180min
-
Measuring Inflation Pressure :
-
Adult : 300 ¡¾15mmHg
-
Neonate : 150 ¡¾15mm Hg
-
Alarm Limit : Adult/Neonate
-
Systolic :
-
High : 1~300/1~150mm Hg
-
Low : 1~300/1~150mm Hg
-
Mean :
-
High : 1~300/1~150mm Hg
-
Low : 1~300/1~150mm Hg
-
Diastolic :
-
High : 1~300/1~150mm Hg
-
Low : 1~300/1~150mm Hg
-
Pulse
-
Measuring : Range 30~254bpm
-
Averaging : 4~16 beats
-
QRS Indication : Audible and Visual
-
Respiration
-
Excitation Current : < 100 uA
-
Excitation Frequency : 50 KHz
-
Measuring Method : Thoracic Impedance Detection
-
Measuring Range : 2~99 bpm
-
Averaging : Every 2 beats
-
Accuracy : ¡¾1bpm
-
Temperature
-
Measuring Range : 15.0~45.0¡Æc
-
Probe Compatible : YSI probe
-
Scale : Celsius Degrees
-
Accuracy : ¡¾0.1¡Æc
-
Print
-
Memorized Data Print
-
Waveform Print
-
Paper : 52mm f40 Thermal Roll Paper
-
Annotations : Time, Date, Vital Sign Digital Readout
-
Display
-
Resolution : 320 (H) 240 (V) Pixels
-
Pixel Pitch : 0.33 0.33mm
-
Viewing Size : 120 (H) 92(V)mm
-
Backlight : CCFL
-
Type : BW STN LCD
-
Trace : 2 Trace (Cascade ECG, ECG+SpO2, ECG+Respiration)
-
Trend : Memorized Waveform
-
Power
-
External Power
-
AC : 85~265V, 50/60Hz
-
DC : 10~28V, 3A
-
Internal Power : NiCd Rechargeable Battery 1.2V, 7ea, 2AH
-
Included Thermal Fuse
-
Operation Time by Internal Power 2 hours typical on full charge
-
Protective Class : Class I
-
Degree of Protection : Type CF
-
Power Consumption : 12WDC
-
Digital In/Output : RS-232C
-
Dimension : 220 mm (W) 160 mm (H) 150 mm (D)
-
Weight : 2Kg Approximately