Novametrix Product Detail
Model: TCO2M
TCO2M is the latest in transcutaneous CO2 and O2 monitoring technology from Novametrix. This lightweight, portable monitor with its easy-to-see display and rugged, reliable sensors, is ideal for use in the NICU, PICU, vascular lab, hyperbaric medicine, wound care, or sleep lab; wherever it is important to provide accurate continuous measurement of transcutaneous CO2 and O2.
Features:
-
Real time graphic trends along with numeric values provide truly continuous CO2 and O2 monitoring.
-
A simple menu system with helpful messages guides the user through setup and alert conditions.
-
Built-in 24-hour trend memory with graphic and histogram displays for analysis of patient history.
-
The lightweight portable monitor can operate from AC power or from its internal battery.
-
Sensor calibration is quick, fully automatic and uses a true two-point method for accurate calibrations.
-
Automatic barometric pressure compensation for accurate calibrations with no need to set/remember "cal values".
-
TCO2M accepts combination CO2/O2 sensors, and CO2 or O2 only sensors for optimum versatility.
-
Sensors provide superior performance, fast response, are rugged, reliable and remembraned in seconds.
-
An automatic site timer and skin friendly attachment systems enhance patient safety and comfort.
Specifications:
-
Measurement Range: Carbon Dioxide 0-255 mmHg (0-25.5 kPA). Oxygen 0-800 mmHg (0-99.9 kPA)
-
CO2 Metabolic Correction: Displayed pCO2 is compensated for metabolic factors related to the temperature effect and CO2 production.
-
Parameter Alerts: CO2 and O2 limit alerts, audible and visible. User selectable, continuously displayed, retained in memory.
-
Alert Silence: Two minute alert silence with LED indicator. Audio Off with LED indicator (user can disable if needed).
-
Temperature Section: User selectable 37-45° C in 0.5° C increments. Can be continuously displayed. Audible & visible alarm if actual/set temperatures differ by >0.2° C. Two fully independent thermistors monitor and control temperature. Heater power shutdown with audible & visible alarm at over-temp (45.5° C) or if thermistors differ by >0.5°.
-
Site Timer: Automatically tracks sensor site duration, provides audible and visible alarms, automatic shutdown of heater power once timer expires. User selectable durations in 0.5 hr increments up to 5, 8, or 12 hour maximum, plus off.
-
Local Power: The power required to maintain sensor temperature is measured and displayed. Range: 0-800 mW. User selectable power reference point allows for quick power trend analysis (LPR).
-
Real Time Trends: Continuously displayed user selectable/scalable real time 10, 20, 30 minute trends of CO2, O2 or Local Power.
-
Trend/Histogram Memory: Battery backed 24 hour trends, 8 second resolution. Trend/Histogram displays 30 min, 2, 4, 8, 12 hours.
-
External Devices: Standard RS232 serial outputs: NovaCARD, computer, printer, others. Optional Analog Module, CO2, O2 and LP.
-
Power: 100-120/200-240 VAC, 50-60 Hz, 40 VA.
-
Battery: Greater than 3 hour operation, 12 hour maximum recharge, lead-acid gel-cell.
-
Display Type: Cold Cathode Display (CCD), 5 × 1.5 inches, W × H (12.7 × 3.8 cm). User adjustable contrast.
-
Monitor Size: 3.3 × 9.0 × 8.0 inches, H × W × D (8.4 × 22.8 × 20 cm). Weight: 8 pounds (3.6 kg).
-
Temperature: Operating 10-40° C (50-104° F), Storage -10-55° C (14-131° F). Relative humidity: 0-90%.
-
Altitude/Pressure: Automatic barometric pressure compensation. Range: 560-800 mmHg (71.9-106.6 kPa).
-
Safety/Regulatory: Designed to comply with EC and TUV (IEC 601-1), UL (2601), and CSA (601-1-M90) standards.
-
Sensor Parameter: Combination CO2/O2, CO2 only or O2 only
-
Sensor Principle of Operation: Stow-Severinghaus pH/Clark-type polargraphic
-
Sensor Response Time (t90): CO2 45 sec, O2 12 sec
-
Sensor Stability: Better than 2 mmHg/hr
-
Diameter/Height: 19/11 mm (0.75/0.41 in)
Request Info
|