Datex-Ohmeda Product Detail
Model: TruSat
The first of its kind — TruSat offers clinicians a real-world oximetry solution.
Features:
- TruSignal Enhanced SpO2 for improved performance during clinical motion and low perfusion
- Same proven durability as in the TuffSat handheld oximeter
- Up to 35 hours of uninterrupted battery life and a fast, full recharge in only 3.5 hours
- Small, lightweight design, only 2.75 lb. (1.24 kg)
- PIr (relative perfusion index measurement) — a quick, easy-to-use, clinical tool for a dynamic numeric reflection of perfusion at the sensor site
- Simple, three-button operation and intuitive menus
- Trend memory option stores up to 48 hours of data, recording a data point every four seconds
- A full, three-year warranty
- Highly visible, backlit screen with large, easy-to-read numerics
- Alarm limits always visible and saved between uses
- Pulse bar waveform
- Color choices include yellow and white
Specifications:
- Measurement
- General
- Pulse oximetry sensors: Datex-Ohmeda OxyTip+ sensors only
- Method: Red and infrared light absorption
- Red LED wavelength range: 650to 670 nm
- Infrared (IR) LED wavelength range: 930 to 950 nm
- Average power: ≤1 mW
- SpO2
- Calibrated for functional oxygen saturation
- Calibration range: 70 to 100%
- Measurement and display range: 1 to 100%
- Display resolution: 1%
- First reading (full accuracy): ≤10 seconds
- Accuracy, Arms (root mean square of paired values; previously represented by ±1 standard deviation):
- 70 to 100% ±2 digits (without motion)
- 70 to 100% ±3 digits (during clinical patient motion1)
- 70 to 100% ±2 digits (during low perfusion)
- Below 70% unspecified
NOTE: SpO2 measurement accuracy is based on deep hypoxia studies using OxyTip+ sensors on volunteer subjects. Arterial blood samples were analyzed simultaneously on multiple CO-oximeters.
NOTE: Accuracy may vary for some sensors; always check the instructions for the sensor.
1 Applicability: OxyTip+ Adult/Pediatric and All Fit sensors
- Pulse rate
- Measurement and display range: 30 to 250 beats per minute (bpm)
- Display resolution: 1 bpm
- First reading (full accuracy): ≤15 seconds
- Accuracy (±1 standard deviation):
- 30 to 250 bpm: ±2 bpm or ±2%, whichever is greater (without motion)
- 30 to 250 bpm: ±5 bpm (during motion)
- 30 to 250 bpm: ±3 bpm (during low perfusion)
- PIr Pulsatile Value
- Measurement and display range: 0.01 to 9.99
- Display resolution: 0.01 PIr
- Monitor
- General
- Lock function: locks/unlocks alarm limits, alarm audio ON/OFF, volume settings, and backlight
- Display
- Liquid crystal display (LCD)
- Pleththysmographic pulse bar (pleth bar): ten-segment column that pulsates to indicate the pulse rate and signal strength (perfusion).
- Display update time:
- SpO2, pulse rate, and PIr values: 1 second ±0.25 second
- Plethysmographic pulse bar: 20 Hz minimum (.05 second)
- Backlight LED: ON or OFF
- Audio
- Pulse rate beep: tone rises as oxygen saturation increases and falls as it decreases
- Adjustable volume: 4-segment on-screen control
- Pulse beep volume: 0 (OFF), 1, 2, 3, or 4 shaded segments
- Alarm volume: 1 (low) , 2, 3, or 4 shaded segments
- Volume intensity at distance of 1 m (3.28 ft): 45 dB minimum to 85 dB maximum
- Alarms
Visual and audible indicators for physiological alarms (SpO2 and pulse rate limit alarms) and technical alarms (sensor condition, button stuck, battery condition, and internal malfunction)
- Alarm limit factory settings:
- High SpO2: 51 to 100% or OFF (—) - OFF
- Low SpO2: 50 to 99% or OFF (—) - 85
- High pulse rate: 30 to 235 bpm or OFF (—) - 130
- Low pulse rate: 30 to 235 bpm or OFF (—) - 40
- Alarm indicator (red/yellow LED):
- High priority alarms: flashing red
- Medium priority alarms: flashing yellow
- Low priority alarms: yellow ON
- Audible alarm tone: pattern varies according to alarm type and priority
- Alarm silence button: 120 seconds or indefinitely
- Alarm silence indicator: screen icon
- AC mains power
- Type: Power supply (AC to DC converter)
- AC power input: 100-240 V, 0.5 A, 50-60 Hz
- Output to monitor: 12 VDC, 1.25 A, 15 watt
- Power indicator (green LED): ON while monitor is connected to AC power
- Line power filter (monitor setting): 50 Hz or 60 Hz
- Battery power
- Type: Internal, rechargeable, nickel metal hydride (NiMH), 3 AH, 12VDC, 150 mA
- Capacity, fully charged, operating at room temperature:
- Without Trend Download option: 35 hours
- With Trend Download option: 24 hours
NOTE: Continuous use of the backlight reduces the time approximately 50%.
- Charging time (full capacity): 3.5 hours typical
- Automatic power off to conserve battery: 20 minutes after monitoring stops
- Battery indicator: 4-segment screen icon; shaded segments represent battery charge (0 = low or depleted; 4 shaded = fully charged)
- Environmental conditions
- Temperature
- Operating: 10 to 40 ºC (50 to 104 ºF)
- Transport and Storage: –40 to 70 ºC (–40 to 158 ºF)
- Relative humidity, noncondensing
- Operating: 20 to 95%
- Transport and Storage: 5 to 95%
- Atmospheric pressure
- Operating: 1060 to 697 hPa
- Transport and Storage: 1060 to 188 hPa
- Approximate elevation
- Operating: –378 to 3048 m (–1240 to 10,000 ft.)
- Transport and Storage: –378 m to 12.2 km (–1240 to 40,000 ft.)
- Dimensions and weight:
- Width/Depth/Height (including handle): 21.8 X 11.5 X 10.3 cm (8.5 X 4.5 X 4 inches)
- Weight: 1.25 kg (2.7 pounds), 1.47 kg (3.2 pounds) with Trend Download option
- Trend download options:
- Trend data storage: 48 hours with a data storage resolution of 1 data point every 4 seconds
- RS-232 serial port:
- DIN 6 circular connector
- 1.92 K baud, 8 data bits, 1 start bit, 1 stop bit
- Handshaking (RTS, CTS), full duplex, no parity
Request Info
|
|