Signal Extraction Pulse Oximeter Product Detail
Model: Rad-8
Compact design. Unmatched clinical performance.
Featuring Masimo SET® pulse oximetry, proven accurate during motion and low perfusion in more than 100 independent and objective studies
The accuracy of Masimo SET pulse oximetry has been shown to reduce false alarms by 95% without missing true clinical events
Simple, easy to use interface for quick setup and alarm management with one touch programming
Large LED color display is easy to read at a distance
Compact, lightweight design is ideal for acute and alternate care settings including long term care facilities, homecare and sleep labs
Features
Sleep Mode easily configures system to perform bedside studies
2 second averaging in sleep mode
Home Mode allows for safe and accurate monitoring and trending at home
RadNet® and RadLink® interface capability for multi-patient remote monitoring
Perfusion Index (PI) indicates arterial pulse signal strength and may be used as a diagnostic tool during low perfusion1
Low Signal IQ® (SIQ) indicator highlights conditions of low signal quality
FastSat™ tracks rapid changes in arterial O2 with high fidelity unlike any other pulse oximeter
APOD™ (Adaptive Probe Off Detection) offers the best probe off detection of Masimo's three sensitivity modes - APOD, Normal and MAX sensitivity
Adjustable averaging 2 to 16 seconds
Nurse call interface
Up to 7 hours of internal battery life when fully charged
72 hours of trending memory
Available in horizontal and vertical configurations
Compatible with Philips Vuelink device interface module
Specifications
Performance Measurement Range
SpO2.......................................................... 1 – 100%
Pulse Rate................................................... 25 – 240 (bpm)
Perfusion Index............................................ 0.02% – 20%
Saturation Accuracy
Saturation...................................................60% to 80%
No Motion2
Adults, Infants, Pediatrics.............................. ±4 digits
Saturation................................................... 70% to 100%
No Motion
Adults, Pediatrics......................................... ±2 digits
Neonate.................................................... ±3 digits
Motion4
Adults, Pediatrics......................................... ±3 digits
Neonate.................................................... ±3 digits
Low Perfusion5
Adults, Pediatrics........................................ ±2 digits
Neonate.................................................... ±3 digits
Pulse Rate Accuracy
Pulse Rate................................................. 25 – 240 bpm
No Motion
Adults, Pediatrics, Neonate.......................... ±3 digits
Motion
Adults, Pediatrics, Neonate.......................... ±5 digits
Low Perfusion
Adults, Pediatrics, Neonate.......................... ±3 digits
Resolution
Saturation (%SpO2).................................... 1%
Pulse Rate (bpm)........................................ 1 bpm
Electrical
AC Power requirements................................ 100-240 VAC, 47-63 Hz
Power consumption..................................... 20 VA Max
Batteries
Handheld
Type......................................................... Sealed lead acid
Capacity..................................................... up to 7 hours6
Charging time............................................. 8 hours
Environmental
Operating Temperature............................... 41°F to 104°F (5°C to 40°C)
Storage Temperature.................................. -40°F to 158°F (-40°C to 70°C)
Operating Humidity..................................... 5% to 95%, non-condensing
Operating Altitude................... 500 mbar to 1060 mbar pressure; -1000 ft to 18,000 ft (-304 m to 5,486 m)
Physical Characteristics
Dimensions..................... 8.2” x 6.0” x 3.0” (20.8 cm x 15.2 cm x 7.6 cm)
Weight.......................... 2.1 lbs=.908 kg=32oz
Modes
Averaging mode7..................... 2, 4, 8,10, 12, 14 or 16 seconds
Sensitivity............................... APOD, Normal and Max8
Alarms
Audible and visual alarms for high and low saturation (1% to 100%), pulse rate (25 - 240 bpm), sensor condition, system failure and low battery
Alarm volume......................... High: 85dB (min) - Low: 45dB (min)
Display/Indicators
Data display............. % SpO2, alarm status, alarm silenced status, AC power, Signal IQ/pleth bar, perfusion index bar, battery status, no sensor, sensor off
Type..................... LED
Compliance
Safety Standard for Medical Equipment........... IEC 60601-1 2nd Edition, UL 60601-1, CAN/CSA C22.2 No. 601-1, JIS 0601-1
Type of Protection........ Class 1 (AC power) Internally powered (battery power)
Degree of Protection-Patient Cable.......... Type BF, Defib Proof-Applied Part
Rad-8 Mode of Operation....................... Continuous
EMC Standard...................................... EN60601-1-2, Class B
Request Info
|
|