Welch Allyn Product Detail
Model: CP 100
A reliable, easy-to-use, and cost-effective ECG device for almost any office setting.
Features:
- A 12-lead resting ECG that provides accurate measurements
- Optional MEANS ECG interpretation software for a second opinion
- High-resolution thermal printer generates full-sized standard and customized reports
- Lead quality graphic saves valuable time by highlighting poor connections
- Full alphanumeric keyboard and LCD screen for easy patient data entry
Specifications:
- Dimensions: 15.6 in (39.7 cm) X 6.2 in (15.7 cm) X 16.2 in (41.1 cm)
- Weight: 11.6 lb (5.3 kg)
- Keyboard type: Elastomer keypad with complete alphanumeric keys
- Paper type: 8.25 x 11 inches (21 x 28 cm), Z-fold thermal paper, 200 sheets
- Thermal printer (internal): Computer-controlled dot array, 8 dots/mm
- Thermal paper speeds: 10, 25, 50 mm/s
- Gain settings:
- Auto ECGs: 5, 10, 20 mm/mV, AUTO
- Rhythm ECGs: 5, 10, 20 mm/mV
- Report print formats:
- Auto ECGs: 3x4, 3x4+ 1R, 3x4 +3R, 6x2, 12x1, 6x2 50 mm/s, 6x2 Ext.
- Rhythm ECGs: 3 leads at a time
- Average cycles: 3x4 50 mm/s + 3R, 6x2 50 mm/s + 6R
- Frequency range: 0.3 to 150 Hz
- Digital sampling rate: > 1,000 samples/second/channel
- Pacemaker detection: ANSI/AAMI EC11
- Power requirement: Universal AC power supply 100-240 V ~, 50/60 Hz ~, 2.2 A maximum
- Fuses:
- AC: Time-delay type, 2 amp 250 V rating, Littlefuse 0215002 or equivalent.
- DC (battery fuse): Fast-acting type, 10 amp 32 V rating, Bussman ATC-10 or equivalent.
- Lead configurations: Standard, Cabrera
- Rechargeable battery: Lead acid gel, 6 volt, 5 AH. Prints up to 100 continuous ECGs per charge. 12-hour recharging.
- Filters: 0.5 Hz high-performance baseline filter. 35 Hz muscle-tremor filter. AC-interference filter 50 Hz or 60 Hz.
- Safety, EMC, and regulatory compliance:
- ANSI/AAMI EC11* UL60601-1
- CAN/CSA C22.2 No. 601.1 IEC/EN 60601-1
- CAN/CSA C22.2 No. 601.1.1 IEC/EN 60601-1-1
- CAN/CSA C22.2 No. 601.1.2 IEC/EN 60601-1-2
- CAN/CSA C22.2 No. 601.1.4 IEC/EN 60601-1-4
- CAN/CSA C22.2 No. 601.2.25 IEC/EN 60601-2-25
- Standard connectivity: SD memory card slot (for use with cards ? 64 MB)
- Connectivity with electronic medical records: Supported through the Welch Allyn CardioPerfect workstation software
- Electrodes: Rigorously tested for conductivity, adhesion, and hypoallergenic qualities, and exceed all AAMI standards
- Cables and wires: Meet or exceed IEC 60601
- Environmental operating conditions:
- Temperature: +10° C to +40° C (+50° F to +104° F)
- Relative humidity: 15 - 95% noncondensing
- Altitude: 700 - 1060 hPa
- Environmental storage conditions:
- Temperature: -20° C to +49° C (-4° F to +120° F)
- Relative humidity: 15 - 95% noncondensing (30 - 70% for printing)
- Altitude: 500 - 1060 hPa
- Protection against electric shock: Class I, internally powered, Type CF
- Protection against ingress of water, per IEC 60529: IPX0
- Mode of operation: Continuous
Request Info
|