LME Infusion Pump Product Detail
Model: Infusion Pump LME-4810
Functions and Features
Three kinds of modes: Rate mode, Time mode, Body Weight mode
High-definition large screen color LCD, making it easy to operate
Double CPU allows for safer and more reliable infusion
Infusion rate from 1ml/h ~ 2,000ml/h
Drop Counter inspection and automatic controls, automatic conversion between ml/hour and drop/minute, one-touch switch display
Display of the drug names
Data lock used to prevent accidental modification of infusion parameters
Display batter capacity
Up and Down pressure sensors prevents free flow
Compatible with most standard brands of infusion tubes
Technical Parameters
Size of Outer Shell: 129 x 150 x 172mm (L x W x H)
Power Supply: AC100V ~ 220V, 50Hz ~ 60Hz
Flow Rate Range: 1ml/h ~ 2,000ml/h
Bolus Rate: 1000ml/h
KVO Rate: Adjustable: 1.0ml/h ~ 5.0ml/h
Volume Limit (initially validity): 0.1ml/h ~ 9999.9ml/h ("0" for turning off the setting function of the volume limit)
Battery Recharge: Rechargeable Lithium Polymer battery, work for 4 hours after being fully recharged (25ml/h)
Compatible: Standard brands of infusion tubes
Occlusion Value: 400mmHG ~ 900mmHG, adjustable, graphical display the pressure
Mode Selection: 1. Rate Mode 2. Time Mode 3. Body Weight Mode
Rate Mode: Rate: 1ml/h ~ 2000.0ml/h
Time Mode: Volume: 0.1ml ~ 999.9 ml Time: 1min ~ 2000min
Body Weight Mode: Weight: 0.1kg ~ 300.0kg Drug: 0.1mg ~ 999.9mg Volume: 0.1ml ~ 999.9ml Dose: 0.1 ~ 9999.9 (restricted to unit, drug, solution, weight and infusion pipe) Unit: mg/kg/h, ug/kg/min
Drop Counter Mode: Adjustable between 1 drop/min ~ 670 drops/min
Alarm Function: Infusion completion, OCCL. (Upper/Down), low voltage, air bubbles, installation error of infusion pipe, door open, abnormal 1, abnormal 2
Panel Instruction Info.: Infusion rate, volume limit, accumulated volume, battery capacity, battery charging indicator, AC power connection indicator, infusion completion, OCCL., air bubbles inside the infusion pipe, bed No., pressure value, flow rate, etc.
Max. Weight: < 2.2kg
Power Consumption: 25VA, 4 hours after fully charged
Infusion Increment: 0.1ml
Accumulated Volume: 0.1ml ~ 9999.9ml
Infusion Accuracy: ±5% (up to ±2% after calibration)
Preset Time Setting (initial validity): 0 ~ 99 hours 59 minutes ("0" for turning off the setting function of pre-set)
Request Info
|