The Flo-Gard 6201 Volumetric Infusion Pump from Baxter can deliver a wide variety of fluids over a broad range of infusion rates.
Features:
Uses only standard Baxter solution administration sets.
Configurable settings allow the pump to be custom-tailored to best suit the hospital's needs.
Computer control capability allows remote control or monitoring of the Flo-Gard 6201 pump via a standard RS232-C interface.
Pumps a wide variety of fluids, including blood and fat emulsions.
Suitable for use in epidural administration.
Safety clamp automatically occludes the tubing when the pump door is opened.
Occlusion sensors detect both upstream and downstream restrictions. Sensitivity of the downstream occlusion sensor is selectable through the pump's configuration.
Flow check display shows resistance to flow.
Ultrasonic air-in-line detector. Sensitivity of the air detector is selectable through the pump's configuration.
The Programmed Delivery Profile (PDP) enables manual programming or ramped calculation of up to 10 sequential infusion programs for situations where multiple flow rates are indicated.
Slide clamp loading option provides flow shutoff when I.V. set is removed from the pump.
Locking control panel helps prevent patient tampering.
An independent secondary medication program. The pump automatically switches over to the primary program upon completion.