Baxter Product Detail
Model: Colleague
Basic Programming Parameters:
- Flow Rate Range (Primary):
- 0.1 – 99.9 ml/h (0.1 ml increments)
- 1 – 1,200 ml/h (1 ml increments)
- Flow Rate Range (Piggyback):
- 0.1 – 99.9 ml/h (0.1 ml increments)
- 1 – 500 ml/h (1 ml increments)
- Volume to be Infused:
- 0.1 – 99.9 ml/h (0.1 ml increments)
- 1 – 9,999 ml/h (1 ml increments)
- KVO Flow Rate: 0.1 – 5 ml/h (0.1 ml increments) (configurable) or programmed rate, whichever is less
Optional Infusion Modes
- Volume-Time Programming
- Piggyback Infusion Program Function
- Piggyback Complete Callback Alert
- Flow Check Display of In-Line Resistance to Flow
- Prime Utility
- Clinician Override of Downstream Occlusion Limits
Configurable Sensor Settings:
- Air Alarm Sensitivity: Measured Accumulations of approximately 25, 50, 100 or 150 microliters
- Nominal Downstream Occlusion Values for Alarm (psig):
- Rate Range (mL/hr): <21 - Minimum: 2, Moderate: 5, Maximum: 9
- Rate Range (mL/hr): 21-200 - Minimum: 4, Moderate: 8, Maximum: 12
- Rate Range (mL/hr): 201-1,200 - Minimum: 6, Moderate: 11, Maximum: 15
Features:
- Label Library: Up to 64 preset drug labels and applications may be selected.
- Dose Calculator: 9 optional dose formulas including: mg/min, mg/hr, mg/kg/min, mg/kg/hr, mcg/min, mcg/hr, mcg/kg/min, mcg/kg/hr, units/hr.
- Volume History: Provides Primary, Piggyback and Total Volume Delivered, current time and date as well as the time and date that Volume History was last cleared.
- Programmable Personalities: Up to 8 unique configuration parameter sets with selectable feature options that can be retained in memory, named and selected on demand.
- Battery Charge Level: Displays a measurement of approximate charge level or remaining battery capacity.
- Auto Restart: Enables the pump to automatically restart when a downstream occlusion has been corrected. This is a configurable option with a range of 0-9 restart occurences before manual intervention is required.
- Auto Tube Load: Mechanical automation of IV set loading and unloading.
- Keyed On/Off Slide Clamp Loading Mechanism: Supports IV set-based free-flow prevention upon set removal from pump.
- Program Memory: Memory retention feature retains infusion settings after the pump is turned off for 5 hours.
Specifications:
- Weight: Approximately 10.4 lbs (4.7 kg) (including mounting clamp)
- Size: Approximately 10.0" H x 7.75" W x 7.75" D (254 x 197 x 197 mm) (without clamp knob or power cord)
- Power Cord: 2.9 m (9 ft) long with hospital grade plug
- AC Power Requirements: 100-120 / 220-240 VAC, 50/60 Hz
- Leakage Current: Less than 300 microamps earth leakage (tested per UL 2601)
- External Fuses: 1.6 amp Type T (time delay) 250V
- Battery Supply System
- Sealed lead acid
- 5 hours operating time @ 100 mL/hr (fully charged new battery)
- 8 hours to recharge to 80% of capacity
- Internal charge system recharges battery whenever connected to AC outlet
- Ambient Operating Temperature: 59°F to 100°F (15°C to 38°C)
- Storage Temperature: -20°F to 135°F (-29°C to 57°C)
- Ground Impedance: Less than 0.2 ohms (tested per UL-2601)
- Vacuum Fluorescent Pump Channel Display
- LCD Screen Contrast Adjustment Control
- Alert / Alarm Audible Adjustment Control
- Alert / Alarm Interval (1-7 seconds)
- Alarm Off Interval (1-7 seconds)
- Calculator-Style Numeric Programming Keypad
- Multi-Functional Mounting Clamp
- Nurse Call Option (Fixed output)
- RS232 Communications Port
- Time Setting (real time entered in hours and minutes)
- Date Setting (month / day / year)
- Diagnostics / Service Functions