LME Product Detail
Model: YP-2000
The YP-2000 infant incubator is the most advanced critical care unit for newborn infants. Air temperature or skin temperature are servo-controlled by micro-computer with oxygen concentration monitoring system and humidity control system and tray (optional); integrated sensor box including the skin temperature sensor, air temperature sensor, oxygen concentration sensor and humidity sensor; the air circulation system can keep the uniformity of bed temperature; double wall hood can be opened on the front and back side, x-ray cassette can be put under the bed; quite operation windows; inclination of infant bed can be adjusted in 0~5°; ,the water tank can be pulled out freely; LED digital display, set-temp, skin temp 1, 2, air temp, oxygen concentration, humidity, heating power, and alarm classification and so on; >37ºC temperature setting; 48hrs memory curve;8 self-checking alarms with audible and visible; RS 232 connector, data output. Electric lift stand, and drawers.
Standard configuration:
Temperature controller, humidity system (optional), oxygen infusion (optional), infant bed, integrated sensor box, x-ray cassette frame, humidity water tank, I.V. Pole, tray, air filter, lift stand.
Specifications:
- Power supply : AC220V, 50HZ
- Power input: ≤1300VA
- Air Temperature control range: 25ºC~37ºC can override to 39ºC
- Skin Temperature control range: 34ºC~37ºC can override to 38ºC
- Sensor precision: ≤0.3ºC
- Changing of temperature: ≤0.5ºC
- Uniformity of temperature: ≤0.8ºC(bed is on the level position)
- Temperature rising time: ≤40min (environmental temperature +22ºC)
- Noise level: ≤55dB (A) (environmental noise ≤40dB)
- Inclination of bassinet: 0°~5°
- Adjustment height of lift stand: < 200 mm
- Indication range of oxygen concentration: 0%~99%
- Setting range of oxygen concentration: 20%~60% (1% each time)
- Indication precision of oxygen concentration: ±3%
- Humidity indication range: 0%RH~90%RH
- Humidity control range: 20%RH~90%RH (1% each time)
- Humidity indication precision: ±6%RH (environmental temperature 20ºC~40ºC~environmental humidity 10%RH~90%RH)
- Indication presicion of humidity: ±6%RH
- Capacity of water tank: 1000 ml
- Working time of water tank when it is filled with water: ≥12 hours (less than 60%RH)
Alarms:
- Audible and visible alarm indication, and LED display can indicate the real temperature (except for power failure alarm and system alarm). For the following alarm 1-6(except for lower deviation), temperature controller can cut off the power supply of heater automatically.
- Over-temp alarm: when the air temperature control ≤38ºC(setting temperature is less than 37ºC ), or ≤40ºC(setting temperature is more than 37ºC): Skin temperature control ≤40ºC
- Deviation alarm: air temperature control ±3.0ºC, skin temperature control ±1.0ºC
- Sensor alarm: when the sensor is disconnected, open circuit, short circuit or skin sensor 1 is not in the right position
- Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working
- Power failure alarm: when there is no power or the power cord is disconnected.
- System alarm: when there is E2ROM malfunction inside of temperature controller.
- Oxygen concentration alarm: ±5%O2
- Humidity deviation alarm:±15%RH
Package:
- Gross Weight: Main body 70Kg / Cabinet 60Kg
- Net Weight: Main body 59Kg / Cabinet 50Kg
- Measurement: Main body 1100mm × 707mm × 888mm
- Cabinet: 1029mm × 697mm × 831mm
Request Info
|
|