Mindray Product Detail
Model: PM9000 Express
Features:
- 12.1" color TFT with maximum 8 waveforms
- Rich in ECG software
- OxyCRG software available
- 40 seconds full-disclosure waveforms review
- Drug dose calculation
- Bed to bed view
- Networking capacity
- Multi-Gas/O2 available
- Multilanguage available Masimo SET SpO2 available
Specifications:
- Safety: IEC60601-1 approved, CE marking according to MDD93/42/EEC
- Dimensions and Weight:
- Dimension: 318mm (W) x 264mm (H) x 152mm (D)
- Weight: 5.4kg
- Operation Environment:
- Power Requirements: AC100-250V, 50/60 Hz
- Temperature: 0-40℃
- Humidity: 30-85%
- Application Scope: Adult, pediatric and neonatal
- Performance Specifications:
- Display: 12.1” color, TFT Resolution: 800x600
- Trace: 8 waveforms
- Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
- Indicator:
- Alarm indicator light
- Power indicator light
- QRS beep and alarm sound
- Interface: Net Port
- Battery: Rechargeable
- Trend time: 1~72 hours
- Alarm: 3-level audible and visual alarm
- Networking: Only connected to Mindray Hypervisor Ⅲ central monitoring system
- Recorder:
- Built-in, thermal array, 2 channels
- Record width: 48mm, (record paper 50mm)
- Record speed: 25mm/s, 50mm/s
- ECG
- 5-lead and 3-lead selectable
- Input: 5-lead: RA; LA; RL; LL; V or R; L; N; F; C 3-lead: RA; LA; LLor R; L; F
- Lead selection: 5-lead: I; II; III; avR; avL; avF; Vx 3-lead: I; II; III
- ECG waveform: 2 channels
- Gain selection: x0.25; x0.5; x1; x2; auto
- Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
- Heart Rate range: Adult: 15-300bpm; Pediatric/Neonatal: 15-350bpm
- Accuracy: ±1bpm or ±1%, whichever is greater
- Resolution: 1bpm
- Protection: Withstand 4000VAC/50Hz voltage in isolation, Against electrosurgical interference and defibrillation
- CMRR:
- Diagnostic mode: >90dB
- Monitoring mode: >105dB
- Operating mode: >105dB
- Calibration signal: 1mV±5%
- Bandwidth: 0.05-100Hz
- Alarm range: Adult:15-300bpm, Pediatric/Neonatal: 15-350bpm
- S-T detection:
- YES
- Measurement range: -2.0mV2.0mV
- Alarm range: -2.0mV2.0mV
- Arrhythmia analysis: YES
- Alarm: YES, audible and visual alarm, alarm events recallable
- Respiration
- Method: RA-LLimpedance
- Measurement range: Adult: 6120rpm; Pediatric/Neonatal: 6150rpm
- Resolution:1 rpm
- Accuracy:±2 rpm
- Alarm range: Adult: 6120rpm Pediatric/Neonatal: 6150rpm
- Apnea alarm: YES
- NIBP
- Method: Oscillometric
- Operation modes: Manual/Automatic/Continuous
- Measurement unit: mmHg/Kpa selectable
- Measurement types: Systolic, Diastolic, Mean
- Measurement range: Adult: 10-270mmHg Pediatric: 10-200mmHg Neonatal: 10-135mmHg
- Over-pressure protection: YES
- Resolution: 1mmHg
- Accuracy: Mean error and standard deviation per ANSI/AAMI SP-10
- Alarm: Systolic, Diastolic, Mean
- SpO2
- Measurement range: 0~100%
- Resolution: 1%
- Accuracy: ±2digit (70-100%); 0-69% unspecified
- Alarm range: 0~100%
- Pulse rate:
- Range: 25250 bpm
- Resolution: 1 bpm
- Accuracy: ±2 bpm or ±2%, whichever is greater
- Alarm range: 25~250 bpm
- Masimo SET SpO2
- Range Saturation (SpO2): 0-100%
- Pulse Rate: 25-240 bpm
- Signal Strength (SS):
- 0.2%: Transmission (dB) > - 50
- 0.02 - 0.2%: Transmission (dB) >= -30 -100 x Signal Strength (%)
- Accuracy:
- Adults
- SpO2 Range: 70%-100% 0% to 69%
- Accuracy (+1Std. Dev.)*: +/- 2 digits Unspecified
- Pulse Rate Range: 25-240 BPM
- Accuracy (+1Std. Dev.)*: +/- 3 BPM
- Neonates
- SpO2 Range: 70%-95% 0% to 69%
- Accuracy (+1Std. Dev.)*: +/- 3 Digits Unspecified
- Pulse Rate Range: 25-240 BPM
- Accuracy (+1Std. Dev.)*: +/- 3 BPM
- Temperature
- Measurement range: 050℃
- Resolution: 0.1℃
- Accuracy: ±0.2℃(including probe)
- Channel: Dual-channel
- Alarm range: 0-50℃
- IBP
- Measurement range: -10-300mmHg
- Channel: 2 channels
- Pressure transducer:
- Sensitivity: 5(µV/V/mmHg)
- Impedance range: 300-3000ohm
- Transducer sites: ART, PA, CVP, RAP, LAP, ICP
- Measurement unit: mmHg/Kpa selectable
- Resolution: 1mmHg
- Accuracy: ±1mmHg or ±2%, whichever is greater
- Alarm range: -10300mmHg
- CO
- Method: Thermodilution
- Measurement range:
- CO: 0.1-20 lit./min.
- TB: 23-43℃
- TI: 0-27℃
- Resolution:
- CO: 0.1 lit./min.
- Temperature: ± 0.1℃
- Accuracy:
- CO: ±0.1Lit/min or ±5%, whichever is greater
- TB: ±0.1℃
- TI: ±0.1℃
- EtCO2
- Method: Infrared Absorption
- Measuring Mode: Sidestream or Mainstream (optional)
- Measurement range:
- EtCO2: 0-99mmHg
- InsCO2: 0-99mmHg
- AwRR: 0-150rpm
- Resolution:
- EtCO2: 1mmHg
- InsCO2: 1mmHg
- AwRR: 1rpm
- Accuracy:
- CO2concentration:
- ±2mmHg (0-40mmHg)
- ±5% of reading (41-76mmHg)
- ±10% of reading (77-99mmHg)
- AwRR: ±2rpm
- Alarm range: Same as Measurement range
Apnea alarm: YES
- Multi-Gas/O2
- Method: Infrared Absorption
- Gas sorts: CO2, N2O, Des, Iso, Enf, Sev, Hal, O2 (optional paramagnetic sensor)
- Measurement range:
- CO2: 0~10% (076mmHg)
- N2O: 0~100%
- O2: 0~100%
- Enf, Iso, Hal: 0~5%
- Sev: 0~8%
- Des: 0~18%
- Data output: Fi and Et values
- Respiration rate: 4~60 ±1bpm
- Other:
- Up to 3 waveforms displayed
- Agent mixture detection
- MAC value displayed
Request Info
|
|
Back |
|
|