Model: PM9000 Express

Features:
  • 12.1" color TFT with maximum 8 waveforms

  • Rich in ECG software

  • OxyCRG software available

  • 40 seconds full-disclosure waveforms review

  • Drug dose calculation

  • Bed to bed view

  • Networking capacity

  • Multi-Gas/O2 available

  • Multilanguage available Masimo SET SpO2 available

Specifications:
  • Safety: IEC60601-1 approved, CE marking according to MDD93/42/EEC

  • Dimensions and Weight:

    • Dimension: 318mm (W) x 264mm (H) x 152mm (D)

    • Weight: 5.4kg

  • Operation Environment:

    • Power Requirements: AC100-250V, 50/60 Hz

    • Temperature: 0-40℃

    • Humidity: 30-85%

  • Application Scope: Adult, pediatric and neonatal

  • Performance Specifications:

    • Display: 12.1” color, TFT Resolution: 800x600

    • Trace: 8 waveforms

    • Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s

    • Indicator:

      • Alarm indicator light

      • Power indicator light

      • QRS beep and alarm sound

    • Interface: Net Port

    • Battery: Rechargeable

    • Trend time: 1~72 hours

    • Alarm: 3-level audible and visual alarm

    • Networking: Only connected to Mindray Hypervisor Ⅲ central monitoring system

    • Recorder:

      • Built-in, thermal array, 2 channels

      • Record width: 48mm, (record paper 50mm)

      • Record speed: 25mm/s, 50mm/s

  • ECG

    • 5-lead and 3-lead selectable

    • Input: 5-lead: RA; LA; RL; LL; V or R; L; N; F; C 3-lead: RA; LA; LLor R; L; F

    • Lead selection: 5-lead: I; II; III; avR; avL; avF; Vx 3-lead: I; II; III

    • ECG waveform: 2 channels

    • Gain selection: x0.25; x0.5; x1; x2; auto

    • Sweep speed: 12.5mm/s, 25mm/s, 50mm/s

    • Heart Rate range: Adult: 15-300bpm; Pediatric/Neonatal: 15-350bpm

    • Accuracy: ±1bpm or ±1%, whichever is greater

    • Resolution: 1bpm

    • Protection: Withstand 4000VAC/50Hz voltage in isolation, Against electrosurgical interference and defibrillation

    • CMRR:

      • Diagnostic mode: >90dB

      • Monitoring mode: >105dB

      • Operating mode: >105dB

    • Calibration signal: 1mV±5%

    • Bandwidth: 0.05-100Hz

    • Alarm range: Adult:15-300bpm, Pediatric/Neonatal: 15-350bpm

    • S-T detection:

      • YES

      • Measurement range: -2.0mV2.0mV

      • Alarm range: -2.0mV2.0mV

    • Arrhythmia analysis: YES

    • Alarm: YES, audible and visual alarm, alarm events recallable

  • Respiration

    • Method: RA-LLimpedance

    • Measurement range: Adult: 6120rpm; Pediatric/Neonatal: 6150rpm

    • Resolution:1 rpm

    • Accuracy:±2 rpm

    • Alarm range: Adult: 6120rpm Pediatric/Neonatal: 6150rpm

    • Apnea alarm: YES

  • NIBP

    • Method: Oscillometric

    • Operation modes: Manual/Automatic/Continuous

    • Measurement unit: mmHg/Kpa selectable

    • Measurement types: Systolic, Diastolic, Mean

    • Measurement range: Adult: 10-270mmHg Pediatric: 10-200mmHg Neonatal: 10-135mmHg

    • Over-pressure protection: YES

    • Resolution: 1mmHg

    • Accuracy: Mean error and standard deviation per ANSI/AAMI SP-10

    • Alarm: Systolic, Diastolic, Mean

  • SpO2

    • Measurement range: 0~100%

    • Resolution: 1%

    • Accuracy: ±2digit (70-100%); 0-69% unspecified

    • Alarm range: 0~100%

    • Pulse rate:

      • Range: 25250 bpm

      • Resolution: 1 bpm

      • Accuracy: ±2 bpm or ±2%, whichever is greater

      • Alarm range: 25~250 bpm

  • Masimo SET SpO2

    • Range Saturation (SpO2): 0-100%

    • Pulse Rate: 25-240 bpm

    • Signal Strength (SS):

      • 0.2%: Transmission (dB) > - 50

      • 0.02 - 0.2%: Transmission (dB) >= -30 -100 x Signal Strength (%)

    • Accuracy:

      • Adults

        • SpO2 Range: 70%-100% 0% to 69%

        • Accuracy (+1Std. Dev.)*: +/- 2 digits Unspecified

        • Pulse Rate Range: 25-240 BPM

        • Accuracy (+1Std. Dev.)*: +/- 3 BPM

      • Neonates

        • SpO2 Range: 70%-95% 0% to 69%

        • Accuracy (+1Std. Dev.)*: +/- 3 Digits Unspecified

        • Pulse Rate Range: 25-240 BPM

        • Accuracy (+1Std. Dev.)*: +/- 3 BPM

  • Temperature

    • Measurement range: 050℃

    • Resolution: 0.1℃

    • Accuracy: ±0.2℃(including probe)

    • Channel: Dual-channel

    • Alarm range: 0-50℃

  • IBP

    • Measurement range: -10-300mmHg

    • Channel: 2 channels

    • Pressure transducer:

      • Sensitivity: 5(µV/V/mmHg)

      • Impedance range: 300-3000ohm

    • Transducer sites: ART, PA, CVP, RAP, LAP, ICP

    • Measurement unit: mmHg/Kpa selectable

    • Resolution: 1mmHg

    • Accuracy: ±1mmHg or ±2%, whichever is greater

    • Alarm range: -10300mmHg

  • CO

    • Method: Thermodilution

    • Measurement range:

      • CO: 0.1-20 lit./min.

      • TB: 23-43℃

      • TI: 0-27℃

    • Resolution:

      • CO: 0.1 lit./min.

      • Temperature: ± 0.1℃

    • Accuracy:

      • CO: ±0.1Lit/min or ±5%, whichever is greater

      • TB: ±0.1℃

      • TI: ±0.1℃

  • EtCO2

    • Method: Infrared Absorption

    • Measuring Mode: Sidestream or Mainstream (optional)

    • Measurement range:

      • EtCO2: 0-99mmHg

      • InsCO2: 0-99mmHg

      • AwRR: 0-150rpm

    • Resolution:

      • EtCO2: 1mmHg

      • InsCO2: 1mmHg

      • AwRR: 1rpm

    • Accuracy:

      • CO2concentration:

        • ±2mmHg (0-40mmHg)

        • ±5% of reading (41-76mmHg)

        • ±10% of reading (77-99mmHg)

        • AwRR: ±2rpm

      • Alarm range: Same as Measurement range

        Apnea alarm: YES

  • Multi-Gas/O2

    • Method: Infrared Absorption

    • Gas sorts: CO2, N2O, Des, Iso, Enf, Sev, Hal, O2 (optional paramagnetic sensor)

    • Measurement range:

      • CO2: 0~10% (076mmHg)

      • N2O: 0~100%

      • O2: 0~100%

      • Enf, Iso, Hal: 0~5%

      • Sev: 0~8%

      • Des: 0~18%

    • Data output: Fi and Et values

    • Respiration rate: 4~60 ±1bpm

    • Other:

      • Up to 3 waveforms displayed

      • Agent mixture detection

      • MAC value displayed