Valleylab Digital Electrosurgical Unit Product Detail
Model: Force 40
Digital Electrosurgical Unit
Unit has digital display for output power and can be used for any electrosurgical procedure.
Performs all basic functions including blend for both cut and bipolar modes
Output Characteristics |
Max. P-P Voltage |
Rated Load (ohms) |
Nominal Power (watts) |
Crest Factor |
Cut |
3300 |
300 |
300 |
2.1 @ 100W |
Blend 1 |
3800 |
300 |
250 |
3.5 @ 100W |
Blend 2 |
4000 |
300 |
200 |
4.0 @ 100W |
Desiccate |
3500 |
300 |
200 |
3.5 @ 100W |
Fulgurate |
6500 |
300 |
150 |
6.0 @ 100W |
Spray |
9000 |
300 |
150 |
8.5 @ 100W |
Std Bipolar |
550 |
100 |
50 |
1.9 @ 20W |
Precise Bipolar |
550 |
100 |
50 |
1.9 @ 20W |
Output Waveform |
Sinusoid |
Duty Cycle |
Recurrence |
Repetition Frequency |
Cut |
500 kHz |
_______ |
_______ |
_______ |
Blend 1 |
500 kHz bursts |
50% |
31.25 kHz |
_______ |
Blend 2 |
500 kHz bursts |
37.5% |
31.25 kHz |
_______ |
Desiccate |
500 kHz bursts |
_______ |
_______ |
250 kHz |
Fulgurate |
500 kHz |
_______ |
_______ |
62.5 kHz |
Spray |
500 kHz |
_______ |
_______ |
31.25 |
Std Bipolar |
500 kHz |
_______ |
_______ |
_______ |
Precise Bipolar |
500 kHz |
_______ |
_______ |
_______ |
Input Power Source |
|
Nominal Voltage |
110-120 Volts RMS or 220-240 Volts RMS |
Regulation Voltage |
90-135 Volts RMS or 180-270 Volts RMS |
Operating Range |
85-140 Volts or 170-280 Volts RMS |
Current |
Idle: 0.4 A, max. or 0.2 A, max. Cut: 7.0 a, max. or 3.5 A, max. Coag: 4.0 A, max. or 2.0 A, max. Bipolar: 2.0 A, max or 1.0 A, max |
REM Contact Quality Monitor |
Measurement frequency: 80 ± 10 kHz. Measurement current: < 10 μA. Acceptable resistance range: single are pad - < 20 ohms, dual area pad – 5-135 ohms |